Calibration
WTX120
A4500-1.1
HBM: public
139
Function/Secondary selection
Key
Menu
With Taring?
Y
Select whether tare functions are disabled or en
abled.
N
Tare functions disabled
Y
Tare functions enabled (default)
For scales that only operate in gross mode, the
tare functions can be disabled here so as to skip
cost-intensive checking of the tared scale during
legal verification.
Onscreen typeplate?
Y
Select whether the
Onscreen typeplate
is dis
played above the scale display:
Y
Display typeplate
N
Hide; the field is left blank
Update Rate
225
Select the update rate for the digital filter
50
…
800 Hz
Setting range
225 Hz
Default setting
If there is vibration around the scale, the proper
ties of the digital filter can be optimized here. The
update rate should be selected as far away as
possible from the vibration frequency and from
the multiple of the vibration frequency.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...