![HBM WTX120 Operating Manual Download Page 138](http://html1.mh-extra.com/html/hbm/wtx120/wtx120_operating-manual_2111754138.webp)
Calibration
136
A4500-1.1
HBM: public
WTX120
11.8
Adaptation
For more information on the keys necessary to access this menu group, go to
the
.
Menu
Key
Function/Secondary selection
Select Group 1-9
Press 5 times for group
5 Adaptation
Motion Window
0.5D
Motion window
Here the window is set in digits, which must con
tain consecutive weight values for the A/D con
verter to signal a stationary weight value.
OFF
Motion control off
3.0D, 2.0D, 1.0D, 0.5D
Window size
MotionCounter
99
Time span for standstill recognition
Here the time span is set in which consecutive
weight values must occur in the motion window
for standstill to be detected. 1 unit = 0.02 sec.
Example: 20 = 0.4 sec
Filter Size
OFF
Filtration level
Here the filtration level of the digital filter is set.
The following settings are possible:
Off
0
1
to
20
Light to heavy filtering.
With a very unstable scale display (such as in the
case of live cattle weighers) heavy filtering is
recommended.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...