Calibration
130
A4500-1.1
HBM: public
WTX120
Information
To save the zero calibration before calibrating the weighing range, you must
exit Setup at this point. Answer the
Save Parameters
prompt with
Y
es. Then
you can call group 2 again to resume the calibration, skipping the zero calibra
tion.
Menu
Key
Function/Secondary selection
Calibrate Load?
Y
Calibrate weighing range
Y
Calibrate range
N
continue to
Load(mV/V)
step
Notice:
You are recommended to use as large a
calibration weight as possible.
Calibr.Weight
999999
Suggestion for calibration weight
(=weighing range)
Clear the suggested weight value and enter any
part load value
Mount load and confirm; continue with calibration
Calibrating...
Measure the load signal
Load:
9999999
Display the 10 times higher weight value resolu
tion (as a check)
Load(mV/V):
1.40000
Display the standardized signal -
see
Calculate Span
Notice:
The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it
by the
Excitation
value (5V or 10V).
Information
The Clr key can be used to clear the value and enter a new one.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...