16
QUICK START GUIDE
Stirrer Configuration:
Stirrer 1
Dosing Type:
D y n a m i c
min Vol:
0.050 mL
max Vol:
0.500 mL
delta E:
20.000 mV
End Point Mode: pH 1EQ point, 1st Der
Recognition Options:
Threshold: 50 mV/mL
Range: NO
Filtered Derivatives: NO
Pre-Titration Volume: 0.000 mL
Pre-Titration Stir Time: 15 Sec
Measurement Mode: Signal Stability
delta E: 1.0 mV
delta t: 2 Sec
t min wait: 2 Sec
t max wait: 15 Sec
Electrode Type: pH
Calculations: Sample Calc. by Volume
Titrant Name: NaOH
Titrant Conc.: 0.1000 M (mol/L)
Analyte Size: 5.000 mL
Analyte Entry: Manual
Maximum Titrant Volume: 20.000 mL
Stirring Speed: 1400 RPM
Potential Range: -2000.0 to 2000.0 mV
Volume/Flow Rate: 25 mL / 50.0 mL/min
Signal Averaging: 1 Reading
Final Result Format: X.XX
M (mol/L) —> M (mol/L)
V mol mol
-*—*—-
L mol
—————
mL L
—*———
1000mL
V = volume dispensed in liters.
0.100 mol/L -> titrant conc.
1.000 mol/mol -> (sample/titrant)
5.000 mL -> sample volume
Nr Volume[ml]
mV
pH
Graphic
Temp[
0
C] Time
0
0.000
235.2
2.857
0.0
A
1 9 . 1 0 0 : 0 0 : 0 0
1
0.050
234.6
2.866
-10.2
A
1 9 . 0 0 0 : 0 0 : 2 1
2
0.100
233.9
2.880
-15.8
A
1 9 . 1 0 0 : 0 0 : 2 7
3
0.200
232.2
2.908
-16.7
A
1 9 . 1 0 0 : 0 0 : 3 9
4
0.390
231.1
2.928
-6.0
A
1 9 . 1 0 0 : 0 0 : 4 5
5
0.590
228.6
2.970
-12.3
A
1 9 . 1 0 0 : 0 1 : 0 4
6
0.790
226.9
3.000
-8.7
A
1 9 . 1 0 0 : 0 1 : 2 0
7
0.990
225.5
3.024
-6.9
A
1 9 . 1 0 0 : 0 1 : 3 7
8
1.190
224.7
3.038
-4.0
A
1 9 . 1 0 0 : 0 1 : 4 3
9
1.390
223.9
3.051
-4.0
A
1 9 . 1 0 0 : 0 1 : 4 9
Summary of Contents for HI 902 Color
Page 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Page 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Page 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Page 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Page 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Page 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Page 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Page 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Page 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Page 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Page 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Page 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Page 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Page 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Page 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Page 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Page 193: ...A5 11 APPENDIX 5 ...
Page 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Page 196: ...1 2 INTRODUCTION ...
Page 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Page 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...