7-13
pH, mV & ISE MODE
Three buffer entry types are available: Automatic, Semiautomatic and Manual Selection (see
section
7.1.2.1 Buffer Entry Type).
To begin calibration:
•
Press
. If the instrument was calibrated before and the calibration was not
cleared, the old calibration can be cleared by pressing
.
Note:
It is very important to clear calibration history when a new electrode is used.
•
Immerse the pH electrode and the temperature probe approximately 4 cm (1.5")
into a buffer solution and stir gently. The temperature probe should be close to the
pH electrode.
•
Select the pH calibration buffer value with
or
.
•
Press
to update the calibration. Once the reading has stabilized, the calibration
buffer will be added to the Calibrated Buffers section.
•
Rinse the pH electrode and the temperature probe, then immerse them into the next
buffer solution and follow the above procedure or press
to exit the procedure.
Notes:
The new calibration points will replace old ones if the difference between them is
+/- 0.2 pH. Buffers used in older calibrations will not have a solid background.
If calibrating with a standard buffer in MTC mode, the pH value and temperature can
be modified by pressing
. The values can be adjusted using the numeric keys.
Press
to save the new values.
buffer solution: one for rinsing the electrode and one for calibration. If you are measuring in
the acidic range, use pH 7.01 or 6.86 as the first buffer and pH 4.01/3.00 or 1.68 as the
second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as the first
buffer and pH 10.01/9.18 or 12.45 as the second buffer.
For extended range measurements (acidic and alkaline), perform a five-point calibration by
selecting five buffers across the entire pH range.
CALIBRATION PROCEDURE
During calibration, the user has a choice of 8 standard buffers: pH 1.68, 3.00, 4.01, 6.86,
7.01, 9.18, 10.01, 12.45 and up to 5 custom buffers.
For accurate measurements it is recommended to perform a five-point calibration. However, at
least a two-point calibration is suggested. For pH titrations the selected buffers should bracket
your endpoint (e.g.: if your endpoint value is at 8.5, use 7.01 and 9.18 for calibration).
Summary of Contents for HI 902 Color
Page 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Page 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Page 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Page 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Page 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Page 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Page 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Page 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Page 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Page 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Page 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Page 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Page 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Page 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Page 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Page 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Page 193: ...A5 11 APPENDIX 5 ...
Page 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Page 196: ...1 2 INTRODUCTION ...
Page 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Page 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...