7-29
pH, mV & ISE MODE
For accurate calibration and to minimize cross-contamination, use two beakers for each
standard solution: one for rinsing the electrode and one for calibration.
Note:
For accurate measurements, add the appropriate ISA (Ionic Strength Adjustment) to
the calibration standards.
CALIBRATION PROCEDURE:
Before calibrating, make sure that the appropriate Electrode Type and concentration unit has
been selected in ISE Setup.
Up to a five point calibration is possible using any combination of five memorized standard
solutions and five custom solutions.
The ISE calibration and measurement can be performed with or without temperature
compensation. If the temperature compensation option is enabled, the isopotential point
of the electrode must be set in ISE Setup.
The current standard will be manually selected by the user from the available standards list.
The list of available standards depends of the Manual Entry setting.
To calibrate the instrument using Manual Entry:
•
Press
from the main screen. If the instrument was calibrated before and the
calibration was not cleared, the old calibration can be cleared by pressing
.
•
Immerse the Ion Selective Electrode and the temperature probe approximately 2 cm
into the lowest concentrated standard solution.
•
Select the concentration with
or
.
•
When the reading has stabilized, press
to update the calibration. The calibration
point value will be added to the Calibrated Standard list.
•
Select
and repeat the procedure with all of the available standards.
Summary of Contents for HI 902 Color
Page 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Page 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Page 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Page 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Page 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Page 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Page 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Page 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Page 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Page 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Page 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Page 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Page 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Page 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Page 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Page 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Page 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Page 193: ...A5 11 APPENDIX 5 ...
Page 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Page 196: ...1 2 INTRODUCTION ...
Page 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Page 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...