Geophysical Survey Systems, Inc.
SIR® 4000
Manual
MN72-433 Rev F
56
Step 2 – Expert Mode Setup
Once a new project is created you will see a split screen with the main data display window on the left,
the O-Scope window in the center, and the Setup menu on the right.
There are four (4) main menus in the Setup Menu: Radar, Process, Output, and System. To get started
collecting data you’ll need to check or adjust settings in the first three (3).
Recall Setup
If you’ve previously saved a setup you can recall it before making any changes. Otherwise when you do
recall the setup it will overwrite any of your changes with those that were previously saved. Setups can
either be uniquely named or you can use one of the sixteen (16) available setup names.
1
Open the System Menu and select Recall Setup.
2
Select your saved setup from the list using either the Control knob or arrow keys. Once the desired
setup is highlighted press either the Control knob or Enter button.
Radar Menu
1
Collect Mode:
Set to Point.
2
Dielectric/Soil Type:
Choose a Dielectric Constant or Soil Type that best represents the area you are
surveying. For a list of Dielectric Constants see Appendix D: Dielectric Values for Common
Materials.
3
Depth/Time Range:
There are two options for setting the depth or time range.
•
Full Range:
With the system in Collect Setup mode and data scrolling across the screen, drag the
antenna over the area. As you do this watch the scan trace displayed in the O-Scope window.
When the scan trace gets to a depth/time where there is more noise, or interference, than signal
strength, you’ll see the scan trace jump from negative to positive and appear craggy. To setup the
system so that you’re collecting the maximum range of data possible you’ll want to see some of
that noise in the scan trace. If no noise is apparent then you should increase your depth or time
range until about 25% of the scan trace is made up of noise. The noise to signal ratio will change
as you move across your survey area. Keeping some noise on the screen will help ensure that you
always collect the full range of quality signal.
•
Target Specific:
With the system in Collect Setup mode and data scrolling across the screen,
drag the antenna over the area and find a target that you want to image. Note its position on the
screen and at what time/depth it occurs. Ideally, it should be halfway down the screen so that you
can image targets above and below it. If it is not properly located, change the range under the
Scan menu.
4
Scans/Unit:
This is the number of scans that will be collected every foot or meter. This number will
affect how the horizontal scale is displayed during data collection.
Note:
If you recall a setup after configuring the GPS, the SIR 4000 will recall a baud rate of 4800.
If this does not match your previously selected Baud Rate, return to the Introduction Screen and reset
your Baud Rate in the GPS Configuration window.
Summary of Contents for SIR 4000
Page 1: ......
Page 4: ......
Page 116: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 110 ...
Page 140: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 134 ...
Page 150: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 144 ...
Page 156: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 150 ...