Geophysical Survey Systems, Inc.
SIR® 4000
Manual
MN72-433 Rev F
148
120-Shallow-Unshielded
120 MHz standard antenna.
Data Collection Mode: Continuous
Range: 200ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 30 MHz
Vertical Low Pass Filter: 240 MHz
Scans per second: 32
Horizontal Smoothing: 5 scans
Transmit Rate: 50KHz
100 High Power
100 MHz antenna with high power transmitter.
Data Collection Mode: Continuous
Range: 500ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 25 MHz
Vertical Low Pass Filter: 200 MHz
Scans per second: 16
Horizontal Smoothing: 5 scans
Transmit Rate: 12 KHz
100 Very High Power
100 MHz antenna with very high power
transmitter.
Data Collection Mode:
Continuous
Range: 500ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 25 MHz
Vertical Low Pass Filter: 200 MHz
Scans per second: 16
Horizontal Smoothing: 5 scans
Transmit Rate: 6 KHz
80 MHz
80 MHz folded bow-tie antenna.
Note:
The 80 MHz antenna is unshielded.
Data Collection Mode: Continuous
Range: 500ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 25 MHz
Vertical Low Pass Filter: 200 MHz
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 50 KHz
MLF 120 cm
Low Frequency antenna 1.2m length.
Note:
The MLF antennas are unshielded.
Data Collection Mode: Point
Range: 250ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 30 MHz
Vertical Low Pass Filter: 160 MHz
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 12 KHz
MLF 240 cm
Low Frequency antenna length 2.4m
Data Collection Mode: Point
Range: 500ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 15 MHz
Vertical Low Pass Filter: 90 MHz
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 12 KHz
Summary of Contents for SIR 4000
Page 1: ......
Page 4: ......
Page 116: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 110 ...
Page 140: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 134 ...
Page 150: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 144 ...
Page 156: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 150 ...