GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 25
15.)Screw 2 screws (2159634, BH M4X10 WHT) to assemble the Arm Cable Cover. Refer to the figure
below.
16.)Place the cables like the figure below.
17.)Screw 1 screw (2306565, BH M4X16 WHT) to fix the Power cable. Refer to the figure below.
Figure 8-50 Assembling the Arm Cable Cover
Figure 8-51 Placing the cables
Figure 8-52 Assembling the power cord bracket