Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 63
8-2-25
APS II Assy
NOTE:
APS Power Supply can not be adjusted.
8-2-25-1
Tools
•
Common philips screwdrivers
•
Nipper
8-2-25-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-25-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove the EMI Cover L. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
3.) Remove HDD power cable.
4.) Extract 3 power connector from back plane.
5.) Disconnect the AC input connector.
6.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
7.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
.
Figure 8-103 APS IIConnectors of backplane
AC Input Connector