GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 4-9 - Board Function Checks
4 - 39
4.) Press “CW” button to start CWD Mode.
5.) Scan in CWD mode (with Virtual Convex enabled).
6.) Press “B” button to scan in “B” mode.
7.) Press Virtual Convex button to disable Virtual Convex if not already disabled.
8.) Press “CW” button to start CWD Mode.
9.) Scan in CWD mode (with Virtual Convex disabled)
10.)Compare the CWD (with Virtual Convex enabled) sensitivity and CWD (with Virtual Convex
disabled).
11.)Pass if two modes have the same sensitivity.
12.)Do the service login. Refer to the 5-7-3-2 " Service login" on page 5-28
13.)Select "Diagnostics" => "LOGIQ P6 Diagnostics" => "FRU Test" => "CWD64" => "CWD64 Overall"
NOTE:
For LP6 PRO, select 'LOGIQ P6 PRO Diagnostics' => 'FRU test' => 'ACWD'
=> "ACWD
Overall"
14.)After All diagnostic function check of CWD64 should be Passed.Refer to the Table 7-5, “FRU Test,
Figure 4-39 CW mode
Figure 4-40 Virtual Convex Disabled.
Summary of Contents for LOGIQ P6
Page 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Page 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Page 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Page 477: ......