Section 8-7 - Mechanical Option Installation instruction
8 - 163
8-7-1-4
Assembly procedure
1.) Insert two of the flexible arm stoppers (A+B) to the Power Cable and DVI cable in the Order
illustrated on the figure below. Make sure that chamfered shape of the flexible Arm stopper should
face inside. Refer to the figure below.
2.) Insert oilless bush into the flexible arm neck pipe and insert two cables into the flexible arm neck
pipe. Refer to the figure below.
3.) Insert the flexible arm into the flexible arm neck pipe. Refer to the figure below.
Figure 8-250 Changing the MON PWR CONN LONG cable
Figure 8-251 Inserting the oilless bush
Figure 8-252 Inserting the flexible arm
Flexible Arm Stopper