Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 93
5.) Disconnect the SATA Power Cable and SATA Cable from HDD Assy after removing 2ea cable ties.
And unscrew 4screws to remove the HDD Fixture bracket.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-43-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-138 Separate HDD and HDD Fixture Bracket
Table 8-44 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-43. Equipment
passed all required tests and is ready for use. “
Cable Ties