8 - 80
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-34
Rear Handle
8-2-34-1
Tools
•
Common philips screwdrivers
•
Allen/Unbraco wrench
8-2-34-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-34-3
Removal procedure
1.) Unscrew 1 screw. Remove the Handle Dumy brkt.
2.) Unscrew 2 screws (1-2) from the back of the rear handle. Refer to the Figure 8-121.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-34-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-120 Handle Dummy Bracket
Figure 8-121 Rear Handle
Table 8-35 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-34. Equipment
passed all required tests and is ready for use. “
Cover Parts Function Validation
(1)
(2)