8 - 8
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-2-3 17" LCDMON OSD Key Assy Replace Procedure
3.) Take the rear cover off using your hands. Refer to the figure below.
4.) Unscrew 2 screws and remove OSD Key Assy from the front cover .Refer to the figure below.
5.) Disconnect the cable from OSD Key assy. Refer to the figure below.
Figure 8-11 Rear Cover
Figure 8-12 Removing the OSD Key Assy
Figure 8-13 Disconnecting the cable