GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 72
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-30
SYSCONML Assy
8-2-30-1
Tools
•
Common philips screwdrivers
8-2-30-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-30-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
.
3.) Disconnect the AEXP cable. Refer to the Figure 8-112.
4.) Remove the SYSCONML Assy in slot #4 in the nest box.
5.) Eject ane ASIR ROM from old SYSCONML with regular screw driver. Refer to the Figure 8-113.
Seperate ASIR ROM form new SYSCONML as same procedure.
Figure 8-112 SYSCONML Assy
Figure 8-113 ASIR location
SYSCONML Assy
AEXP cable