GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
7 - 26
Section 7-8 - LED Descriptions
7-8-5
APS II Assy
The APS Assy is located at the No. 5 slot of the NEST Assy.
Table 7-14
LEDs on APS IIAssy
LED
Descriptions
Normal State
DP5 (Top)
Digital 5V output indicator
Normally On
LP5
Live Power 5V output indicator
Normally On
AP6
Analog 6V output indicator
Normally On
DP33
Digital 3.3V output indicator
Normally On
SHV+
SHV+ output indicator status indicator
Normally On
HVL+
HVL+ level indicator
Normally On in CFM mode or PW mode
HVH+ (Bottom)
HVH+ level indicator
Normally On in B mode
Summary of Contents for LOGIQ P6
Page 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Page 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Page 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Page 477: ......