Section 7-9 - Trouble Shooting Tree
7 - 41
7-9-11 Network Troubleshooting
Figure 7-25 Network Troubleshooting
No Connection to the Network
Connect the cable and verify that it is
well seated in both ends.
YES
NO
Is the
TCP/IP Cable between
LOGIQ™ P6/P6 Pro and
network outlet connected
and well
Replace the AEXP cable between
SYSCONML and P6 ARP assy
YES
Is the connection OK now?
NO
Connect a Crossed Network Cable
between the Scanner and your PC/
Laptop.
Try to ping from the PC/Laptop to
LOGIQ™ P6/P6 Pro.
If OK, the hardware connection
inside LOGIQ™ P6/P6 Pro is OK.
Try a network cable that is known to
be OK
NO
Is the cable between
SYSCONML asyand P6
APR Assy OK?
YES
Replace the P6 ARP assyl.