Section 4-9 - Board Function Checks
4-9-5
APS II Assy PCB Function Validation Procedure
1.) Turn on the system.
2.) Do the Service Login. Refer to the
5-7-3-2 "Service Login" on page 5-25
3.) Select 'Diagnostics' => 'LOGIQ P6 Diagnostics' => 'FRU test' => 'APS II' => 'APS II Overall'.
4.) Execute the diagnostic test.
5.) All diagnostic test item of APS II should be passed.
6.) Shut down the system.
7.) Restart the system
Figure 4-43 APS II Diagnostic Test