GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 158
Section 8-7 - Mechanical Option Installation instruction
8.) Unscrew 2 screws (2159632, BH M4x6 WHT) to remove the Cam cover L & R. Refer to the figure
below.
9.) Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover. Refer to the figure
below.
10.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Lock cover. Refer to the figure below.
Figure 8-238 Removing the Cam cover L & Cam cover R
Figure 8-239 Removing the Arm bottom cover
Figure 8-240 Removing the Lock cover
Summary of Contents for LOGIQ P6
Page 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Page 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Page 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Page 477: ......