Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 75
8-2-31
SOM Replacement
8-2-31-1
Tools
•
Common philips screwdrivers
8-2-31-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-31-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
.
3.) Remove the SYSCOMML Assy. Refer to the
8-2-30 "SYSCONML Assy" on page 8-72
.
4.) Unscrews 4 screws. Refer to the figure below.
5.) Remove the SOM Assy. Refer to the figure below.
Figure 8-116 4 screws for SOM on SYSCON
Figure 8-117 Remove a SOM from SYSCONML Assy