Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 51
8-2-16
Rear Cover
8-2-16-1
Tools
•
Common philips screwdrivers
8-2-16-2
Preparations
•
Shut down the system, and switch off the main breaker.
8-2-16-3
Removal procedure
1.) Remove the Left and Right side covers. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
2.) Remove 4 rubber caps and the 4 screws from the rear cover. Refer to the figure below.(.
3.) Remove the Rear Cover.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-16-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-91 4 rubber caps and 4screws on Rear Cover
Table 8-17 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-16. Equipment
passed all required tests and is ready for use. “
Summary of Contents for LOGIQ P6