GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 8-4 - Software Installation
8 - 139
8-4-4-10
Product Type Check of the System
To verify that the correct version of Application software was installed, check the
Product Type
of the
Application software. Verification steps are listed below.
1.) Turn on the system. Wait until the screen comes up.
2.) Press the
Utility
key and click
Admin
tab.
3.) Check the value of
Product
item.
4.) The following table is the matching table.
Figure 8-209 Checking Product Type
Installed Application
Value of Product item
P6
Radiology.Arirang2
P6Pro
Radiology.Arirang2.Pro
Table 8-53 Product Type Matching Table