8-2-24
EMI Cover L and R
8-2-24-1
Tools
•
Common philips screwdrivers
8-2-24-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-24-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
2.) Unscrew the 6 screws on left EML cover and 9 screws on thr right EMI cover. Refer to the Figure
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-24-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-102 EMI Cover Land R
Table 8-25 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-24. Equipment
passed all required tests and is ready for use. “