Cardiology
10
-
96
LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Continuous Capture mode
Continuous Capture mode enables the user to perform
acquisition continuously for all views at any level depending on
the selected template configuration. Continuous Capture
consists of temporary saving images acquired in a storage
buffer. To enable best possible use of the limited storage buffer
capacity, a Pause/Capture mode is provided, as opposed to the
normal Freeze/Scan mode. The Pause mode enables scanning
and live display on the screen, without any capture, thereby
leaving the buffer available.
To run Continuous Capture, the user has to select a template
where this feature is activated.
The buffer bar
When entering a level with Continuous Capture enabled, a
buffer bar displays in the window.
The Buffer bar displays the following information:
•
The scanning state
•
Pause (live scanning without storing)
•
Capture (live scanning with storing to buffer)
•
The percentage of the buffer that is filled
•
The buffer filling progression showed by a filling gage
•
The capturing sessions, reflected by the red lines along the
buffer bar
Figure 10-12. Buffer Bar
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...