Using 3D
LOGIQ 7 Online Help
5
-
113
Direction 2392536-100 Rev. 1
3D Acquisition
Acquiring a 3D Scan
To acquire a 3D scan,
1. Optimize the B-Mode image. Ensure even gel coverage.
2. Press the 3D control panel key. Two screens appear.
NOTE:
Set appropriate values for Acq Mode and Scan Plane. Also,
set the scan distance before scanning.
3. To start acquiring the image, press ‘L’ (the left split screen
key).
4. To perform a parallel scan, scan evenly. To perform a sweep
(fan) scan, rock the probe once. Note the distance of the
scan.
5. The 3D volume of interest (VOI) is dynamically assembled
on the right side of the screen.
NOTE:
If the image stops before you’re done scanning, start
acquiring the 3D volume of interest again.
6. To complete the 3D scan, press ‘R’ (the right split screen
key).
NOTE:
You can also press Freeze, but then you need to also press
the 3D key to obtain the final render.
3D Notes
Adjust the 3D dataset brightness with B-Mode or Color Flow
Mode Gain.
Use Colorize to change the color of the active dataset.
Use Zoom to increase the zoom factor of the active dataset.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...