Pediatrics Calculations
LOGIQ 7 Basic User Manual
13
-
9
Direction 2392206-100 Rev. 1
B-Mode Measurements (continued)
d:D Ratio
Measurement
The d:D Ratio measurement measures the percentage of the
femoral head coverage under the bony roof. To make this
measurement:
Measure
From the Touch Panel, select either the right or left side (orientation) and then
select
d:D Ratio
.
A horizontal dotted line displays.
Use the
Trackball
to place the baseline along the ilium. Position the crosshairs
edge at the osseous convexity of the ilium.
Use the
Ellipse
control or
Hip Rotate
to adjust or change inclination.
Set
Press
Set
to fix the baseline.
The system displays a circle representing the femoral head. Use the
Trackball
to position the circle.
Use the
Ellipse
control to size the femoral head circumference.
Set
Press
Set
to fix the femoral head circumference.
The system displays the d:D ratio for the femoral head in the Results Window.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...