Trace Sensitivity
Description
Adjust the trace to follow the waveform for signal strength.
Adjusting
Press the left or right arrow key to select value.
Values
From -10 to 10.
Benefits
If the signal is very faint, changing the Trace Sensitivity will allow
the system to lock onto that signal strength.
PW/CF Ratio
PW/CF Ratio
The PW/CF Ratio is active when “Depend Triplex” is on in triplex
mode. It is used to set the PRF ratio between PW and CFM.
Adjusting
Press
PW/CF Ratio
to select the ratio.
NOTE:
Set “Dependent Triplex” to the on position at the Utility ->
Imaging -> PW menu before using this feature.
Values
1–CFM and PW PRF are the same
2–PW PRF is twice the CFM
4–PW PRF is four time the CFM
Benefits
Optimizes PW PRF in Triplex mode. Decrease aliasing without
changing PRF or CFM.
Duplex
Description
Duplex On: simultaneous B-Mode + PW Mode.
Duplex Off: toggle live B-Mode + PW Mode via B Pause.
Adjusting
Press
Duplex
.
Values
On/Off.
Benefits
Allows you to have both B-Mode and PW Doppler Mode active
at the same time.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...