Quality Assurance
LOGIQ 7 Online Help
18
-
39
Direction 2392536-100 Rev. 1
Gray Scale photography
Description
Poor photography will cause loss of low level echoes and the
lack of contrast between large amplitude echoes. See
Figure 18-11 for more information.
Benefit
When photographic controls and film processors are properly
adjusted, weak echoes, as well as strong echoes, are accurately
recorded on film.
Procedure
1. Adjust the camera according to the manufacturer's
instructions until the hard copy and video display are equal.
2. Scan the phantom and it's echogenic contrast targets.
3. Make a hard copy photograph of the display and compare it
to the image on the video monitor for contrast and weak
echo display.
4. Document all results for future reference and comparison.
Contact a Service Engineer if camera cannot duplicate what is
on the image monitor.
NOTE:
Optimization of brightness/contrast controls on the display
monitor is imperative in order to make sure that the hardcopy
and monitor look alike.
The display monitor is adjusted first. The hardcopy camera or
printer is adjusted to match the display monitor.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...