User Maintenance
18
-
40
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Setting up a Record Keeping System
Preparation
The following is needed:
• Quality
Assurance
binder.
•
Hard copy or electronic file of images.
• Quality
Assurance
Checklists.
•
Display the following information while testing quality
assurance:
• Acoustic
Output
• Gain
• Depth
• Probe
• Dynamic
Range
•
Set up new patient to be the name of the test.
•
Annotate the following:
•
Any control where its value is
NOT
displayed.
• Significant
phantom information.
Record Keeping
Complete the following:
1. Fill out the Ultrasound Quality Assurance Checklist for each
probe, as scheduled.
2. Make a hard copy or archive the image.
3. Compare images to baseline images and acceptable
values.
4. Evaluate trends over previous test periods.
5. File hard copy or electronic file of images and checklist in
Quality Assurance binder.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...