Cardiology Measurements
LOGIQ 7 Basic User Manual
10
-
85
Direction 2392206-100 Rev. 1
Flow Volume (FV)
Flow Volume estimates the volume of blood that flows through a
vessel per unit time. It is derived from a vessel's cross-sectional
diameter obtained from the B-Mode portion of the image and the
mean velocity of flow in the vessel obtained from the Doppler
portion of the image. It is measured in milliliters. When the FV
measurement is made, FVO is automatically calculated.
To measure flow volume:
1. Select
FV
from either the B-Mode or Doppler Touch Panel.
2. Place the dotted horizontal line caliper at each of the
timebase on the Doppler spectrum.
•
If Trace Auto is selected, the waveform is automatically
traced.
•
If Trace Auto is not selected, manually trace the desired
portion of the waveform.
The caliper moves to the B-Mode area.
3. Use the Ellipse or Trace method to measure the
circumference and area of the vessel.
The flow volume (FV) is calculated and displayed in
milliliters. The flow volume output (FVO) is also calculated
and displayed in milliliters/minute.
Flow Volume Output (FVO)
This measurement is used to measure the flow volume output in
a vessel on the Doppler spectrum. It is measured in milliliters/
minute. When the FVO measurement is made, FV is
automatically calculated.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...