Quality Assurance
LOGIQ 7 Online Help
18
-
29
Direction 2392536-100 Rev. 1
Baselines
An absolute necessity for a quality assurance program is
establishing baselines for each test or check. Baselines are
established after the system has been verified to be working
properly at installation or after a repair. If a probe or major
assembly is replaced, new baselines should be generated.
Baselines can be made by adjusting system parameters to
prescribed levels or to the best possible image. The key factor to
remember is reproducibility. The same conditions must be
reproduced for each periodic check.
All system parameters not displayed on the monitor should be
recorded for the permanent record.
Periodic Checks
Periodic checks should be performed in accordance with your
facility’s quality assurance requirements. For the data to be
valid, periodic checks should mimic the baseline setup
parameters.
The resulting image, when scanning the phantom exactly as
before, should be recorded and compared to the baseline.
When a matching image is obtained, it can be assumed that the
system performance has not degraded from the baseline.
If a significant difference between the baseline and periodic
check is noted, double check the system setup and repeat the
test. If the difference between the baseline and periodic check
persists, contact a local Service Representative.
Failing to reproduce the control settings as in the baselines will
introduce errors in the data and potentially invalidate the results.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...