Optimizing the Image
5
-
74
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Gain
Description
Gain amplifies the overall strength of echoes processed in the
Color Flow window or spectral Doppler timeline.
Adjusting
Gain values change depending on the probe and application;
they are not associated with a particular position of the button.
To decrease/increase Gain, adjust the dial to the left/right.
Values
Values vary by probe, application, and multi frequency setting.
Gain displays as dB. Gain values are returned to the factory or
user preset value when you change the following: Probe, Exam
Category, Exam Calcs, New Patient, or Multi Frequency.
Benefits
Allows you to control the amount of color within a vessel or to fill
in or clean out spectral information.
Bioeffects
Gain has no affect on Power Output. However, with increased
Gain, the output level can usually be reduced to produce an
equivalent image quality.
PRF (Pulse Repetition Frequency)
Description
Raises/lowers the PRF on the color bar.
Adjusting
To raise/lower the velocity scale, press PRF/Wall Filter until you
reach PRF, then adjust PRF left/right.
Values
Velocity range is in kHz.
PRF values are returned to the factory or user preset value
when you change the following: Probe, Exam Category, Exam
Calcs, or New Patient.
Benefits
Imaging of higher velocity flow requires increased scale values
to avoid aliasing.
Affect on other
controls
Changing the PRF may affect Power Output, frame rate, and
wall filter. When you adjust the velocity scale, CINE memory is
cleared.
Bioeffects
Changing the PRF range may change the TI and/or MI. Observe
the output display for possible effects.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...