Optimizing the Image
5
-
78
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Color Flow Line Density
Description
Optimizes the Color Flow frame rate or spatial resolution for the
best possible color image.
Adjusting
To adjust the line density, press
Line Density
Touch Panel key.
Values
The current frame rate is shown on the Touch Panel. Frame
Rate/Resolution values vary by: Probe, Exam Category, Exam
Calcs, New Patient, and Frequency.
Settings are returned to factory or user preset value when you
change any of the above.
Benefits
Fetal heartbeat, adult cardiac applications, and clinical
Radiology applications require significantly higher frame rates.
High resolution is useful in situations where very small vessels
are being imaged, e.g., thyroid, testicles.
Affect on other
controls
Frame rate changes the vector density and frame rate.
Bioeffects
Activating color flow frame rate/high resolution may change the
TI and/or MI. Observe the output display for possible effects.
Map
Description
Allows you to select a specific color map. After you have made
your selection, the color bar displays the resultant map.
Adjusting
After you activate Color Flow, the Color Flow Touch Panel
displays. To cycle through available maps, select
Map
, move the
Trackball
to view available maps, and press
Set
to select.
Values
Velocity Maps (VL). Flow shown as blue away/red toward the
probe.
Velocity Variance Maps (VV). Provides a measure of turbulence
(stenosis). Adds green to velocity maps.
Benefits
Shows the direction of the flow and highlights the higher velocity
flows.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...