Vascular Worksheet
LOGIQ 7 Basic User Manual
11
-
27
Direction 2392206-100 Rev. 1
To edit a worksheet (continued)
To delete data:
Delete Value
key is used to delete value (each measured
value). Whenever particular value is deleted, the group (set) of
the measurement/study value is also deleted from the
worksheet as a group (set). When PS value is deleted, ED and
RI is also deleted if this value was measured as the set of RI
measurement.
1. Select
Worksheet
from any page of the Vascular
Calculation Touch Panel Menu.
2. To position the cursor at the field you want to delete or
exclude, move the
Trackball
.
The field is highlighted.
3. Select
Delete Value
.
For Example:
1. If the user measured RI 4 times, however, latest 3 set of
measurements were displayed in the worksheet.
1. Then, the user deleted PS value of #3 from the worksheet.
2. The whole set of #3 measurements is deleted from the
worksheet and #1 set of measurements is shifted and
displayed as below.
Result Number
#2
#3
#4
PS
0.500
0.600
0.700
ED
0.100
0.200
0.300
RI
0.800
0.667
0.571
Result Number
#1
#2
#4
PS
0.400
0.500
0.700
ED
0.000
0.100
0.300
RI
1.000
0.800
0.571
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...