OB/GYN
9
-
36
LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
To select OB/GYN vessel measurements
OB/GYN Vessel measurements use the auto sequence feature.
With this feature, when you select a folder for the vessel you
want to measure, the system automatically starts the first
measurement. It then continues with each of the other
measurements in that study.
1. Select the folder for the vessel you want to measure.
The system shows all the measurements for that vessel.
The caliper for the first measurement is automatically
displayed.
2. Make the measurement.
After you complete each measurement, the system starts
the next measurement. After the last measurement is
complete, the system returns to the OB/GYN Vessel Touch
Panel.
Your system is set up to show the measurements that you
usually make for each vessel. To make a measurement that is
not shown for the selected vessel:
1. Select the folder for the vessel you want to measure.
2. Select Show All.
The system display all possible vessel measurements.
3. Select the desired measurement.
The following pages describe the steps to make each specific
measurement in the OB/GYN Vessel study.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...