4:13
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
Service Technicians Guide
The assignment of concentrates to the alterna-
tives is open - any combination of concentrate
types can be used.
All values which are going to be set are specified
at the label on the concentrate containers.
NOTE! If the values are present as both
concentrated and diluted concentrate, use
the concentrated values.
Example
Following example will show how to preset the
AK 95 S for Gambro BiCart together with A-
concentrate 201.
Bicarbonate concentrates
9
1. Press
.
Bicarbonate alt. 1
9-0
2. Press
.
Liquid Bicarbonate 1
9-0-0
BiCart 1
9-0-1
3. Select
BiCart 1
and press
.
BiCart 1
[gram/l,
molecule]
9-0-1-0
BiCart 1
[mmol/l,
molecule]
9-0-1-1
BiCart 1
[mmol/l,
ion]
9-0-1-2
4. Select one of these alternatives and preset the
concentrates according to the label on the con-
centrate container.
For example:
BiCart 1 [gram/l,
molecule]
and press
.
How to preset concentrate
Enter Service maintenance menu:
MAINTENANCE SUPPORT SELECTION
Logging
Service/Preset
Select
Service/Preset
MAINTENANCE SELECTION
Fch
Preset
Service
Select
Preset
and following text is displayed:
INITIALIZING PRESET MODE
SELECT PRESET TYPE
Manual
Ext Default
Select
Manual
.
Scroll with the
SET-knob
:
Group 8 for presetting Acetate concentrates and
group 9 for Bicarbonate concentrates.
Press
.
• 0 Start up
• 1 Configuration
• 2 Set values and limits
• 3 Override times
• 4 Auto priming
• 5 Rinse Back
• 6 Disinfection
• 7 Single needle
• 8 Acetate concentrates
• 9 Bicarbonate concentrates
AK 95 S has six memories, called “alternatives”
for concentrate presets - three for Acetate mode,
and three for Bicarbonate mode.
The
default
concentrates are:
Acetate mode:
Alt 1: D01
Alt 2: D94
Alt 3: D53
Bicarbonate mode:
Alt 1: 201 + BiCart
Alt 2: 203 + BiCart
Alt 3: D360 + D300 (Liquid bicarbonate)
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...