Performance and speci
fi
cation - Control
system
Blood
fl
ow control
Values for the blood pump(s) are based on a pressure of -150 mmHg
before the arterial blood pump with a pump segment of 7.9 mm. For
pediatric blood tubes with pump segment of 4.0 mm it is also possible
to set blood
fl
ow to 5, 10 or 15 ml/min.
Double needle
Flow rate:
0 and 20 - 500 ml/min (-150 mmHg
prepump pressure, pump segment
diameter 7.9 mm)
Flow accuracy:
±10 ml/min or ±15%, whichever is largest
Accumulated blood vol-
ume:
0 - 327 litres
Volume accuracy:
±0.6 l treatment time (h) or ±15%
Single needle
Flow rate:
0 and 20 - 500 ml/min (-150 mmHg
prepump pressure, pump segment
diameter 7.9 mm)
Flow accuracy:
±10 ml/min or ±18%, whichever is largest
Time control:
2 - 20 seconds (±1 sec.), arterial and
venous time. The actual clamp opening
time is depending on blood
fl
ow rate,
size of expansion chamber and pressure
limits.
Pressure control:
10 - 500 mmHg (±50 mmHg), venous
pressure control
Accumulated blood vol-
ume:
0 to 327 litres
Volume accuracy:
±0.6 l treatment time (h) or ±18%
Heparin administration
Heparinization:
0 to 10 ml/h (±1 ml/5 h or ±5%) The
accuracy is based on tests with 20 ml and
30 ml syringes
Syringe size:
Syringes shall comply with ISO 7886-2.
Upon delivery, the machine is set for a
20 ml Terumo syringe. Syringes of other
volumes and brands can be set.
Stop time:
Heparinization stops before end of
treatment 0.00 to 9.59 h
Counter pressure:
Maximum 400 mmHg
Accumulated heparin
volume:
0 to 999.9 ml
HCEN9369 Revision.10.2005
Program version 12.xx
AK 95 S
Service
m anual - Technical data and spec i
fi
cations
3
:2
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...