4:5
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
Service Technicians Guide
Setting parameters
1. Press the
and the text on the display will
flash.
2. Change the value with the Set-knob.
The “Hold-button”
is used to maintain the
text and values showed on the information
display. When you want to leave a hold position
and turn off the information display, press this
button again.
5. Display and setting group
The information display and its two buttons
together with the Set-knob is used for displaying
and setting all parameters exept the blood flow.
The blood flow is always directly controlled with
the Blood flow-knob.
5.1. Information display
5.2. Blood flow knob
5.3. Set knob
Set
Blood
flow
"
Hold
". If pressed when
the information display is
lit, the display is "held"
forever. If pressed again,
the display disappears.
Information display
- goes blank after
15 seconds, unless "Hold" is pressed.
"
Select
" - selects
changeable values in
the information display
Set-knob
; changes
a selected value
Blood flow-knob
;
changes the blood flow
Main treatment parameters
Anytime
the information display is
dark
, the
machine will display the main treatment param-
eters: Blood flow, Venous pressure, TMP, and
UF volume, when “Select” is pressed:
If “Select” is pressed again, the dialysis fluid
flow will be displayed.
Bl:ml/min V :mmH TMP:mmH UFV:l
0 0 0 0.00
p
g
g
Program version handling
At start-up the program decides which PROM
has the latest status and reads out the valid
PROM-configurations. This will prevent the
possibility of mixing PROMS from different
program versions when installing/updating the
AK 95 S. If the configuration is invalid, an error
code will be issued.
The current program version will be displayed
for a few seconds at start-up.
Night light
All buttons and bar graphs become unlit if
“Mute” button is pressed for three seconds.
Any alarm or operator display activity will
deactivate the function. To activate “Night-light”
the preset-parameter 1-34 has to be set to true.
Fluid flow: ml/min
0
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...