AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
Calibrations
1.
Start the AK 95 S, enter the
Service
menu and select
Calibrations
.
2.
Select
Transducer
,
External
and
PRIMING DETECTOR from the selection
diagram.
3.
Press
to start calibration routine.
4.
The display shows:
INSERT TUBE
Press SELECT to confirm
5.
Insert an empty tube in the priming
detector and press
.
Use the normal tube from the blood line.
6.
The calibration is now done automatically
and the calibration factor is writen into
the EEPROM.
7.
When the info. display shows “EEPROM
STORAGE COMPLETED”, the calibration
is completed.
8.
Press
to exit.
If an technical error is detected, the calibration
must be restarted, or the priming detector
replaced.
Occlusion adjustment, blood pump/
infusion pump
Check surface on the roller at the pump rotor, if
it is damaged exchange the unit. The roller unit
should easily go back to upper end position when
it has been pressed down to end position, ex-
change unit if it gets stuck in any position.
Note!
If the blood lines/infusion lines have a
different wall thickness, the occlusion of the
blood pump must be adjusted to the right size.
Use the gauge pins (stop/go) available in the kit
K40158001. The kit includes following three
different gauge pins:
T = 1.1 (Stop pin = 1.1*2*0.7 = 1.5mm)
T = 1.6 (Stop pin = 1.6*2*0.7 = 2.2mm)
T = 2.0 (Stop pin = 2.0*2*0.7 = 2.8mm)
T = wall thickness of the blood tubing
Use following formula to calculate which gage
pin to use:
2 * wall thickness * 0,7 (mm)
Check the range between the blood pump rollers
and the path according to the picture below.
Adjust range between the blood pump rollers and
the path if it is necessary.
The GO pin should barely pass within the cali-
brating area.
The STOP pin is
NOT
allowed to pass in any
point within the calibrating area.
Note!
Put grease on the GO and STOP pins
when not used and store gage in plastic bag to
prevent it from corroding.
Priming Detector calibration
= Measuring
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...