Drain:
The drain tube outlet must be placed
between
fl
oor level and maximum 1.2
m above the outlet connection from the
fl
uid monitor. An air gap to atmospheric
pressure must always be arranged at the
tube outlet.
Length of drain tube: <
10
m
High pressure guard
transducer:
±500 mmHg, (±10 mmHg or ±5%,
whichever is largest)
Inlet pressure switch
(INPS):
Increasing: +150 mmHg, ±20 mmHg
Decreasing: +99 mmHg, ±20 mmHg
Safety guard switch
(SAGS):
Increasing: -59 mmHg, ±7 mmHg
Decreasing: -74 mmHg, ±7 mmHg
Power supply
Mains voltage:
100 V (at 50 Hz or 60 Hz) , 110 V (at 50
Hz or 60 Hz) or 230 V (at 50 Hz or 60
Hz) (
∼
AC), (±12%). The voltage has to
be speci
fi
ed before installation.
Note:
The user must verify the quality of the protective earth in the
installation.
Power consumption:
2025 W at 230 V 1575 W at 110 V 1275
W at 100 V
Cable length:
3 m
Cable area:
AWG 14 at 100 V and 110 V 1 mm
2
at
230 V
External fuses:
For 100/110 V, 2 x 12 AT for heater. For
230 V, 2 x 12 AT for heater.
Mains plug:
For 100/110 V, Hospital grade, earth
plug, type IEC 83 A5-15 For 230 V, Dual
earth system plug, type IEC 83 C4
Maximum earth leakage
current:
100 µA (100 V) 125 µA (110 V) 200 µA
(230 V)
Maximum patient leakage
current:
10 µA (100 V) 10 µA (110 V) 10 µA
(230 V)
Note:
All leakage currents speci
fi
ed are without external equipment
connected to the AK 95 S.
HCEN9369 Revision.10.2005
Program version 12.xx
AK 95 S
Service
m anual - Technical data and spec i
fi
cations
3
:8
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...