Dialysis
fl
uid preparation
Pressure regulators:
After pressure regulator PR 1 and the
heat exchangers:
80 kPa (0.8 ±0.1 bar)
After pressure regulator PR 2:
130 ±10 mmHg
Temperature:
Adjustable 33 to 40 °C. Accuracy is
valid only if dialysis
fl
uid temperature is
greater or equal to ambient temperature.
Alarm limits:
Adjustable 33 to 40 °C
Accuracy:
+0.5/-1.5 °C (+1,0/-2.5 °C with UFD-kit)
at the dialysis
fl
uid outlet from the
machine.
Heater capacity:
1000 W (+10% / -5%) at 100 V 1300 W
(+10% / -5%) at 110 V 1750 W (+10% /
-5%) at 230 V
Overheat protection:
Reg-temp 80 °C or Cond A-temp 70 °C
in treatment, software Reg-temp 99 °C
in disinfection, software. The heater can
only be on if there is a
fl
ow through the
heater. The
fl
ow is detected by a
fl
ow
switch.
Flow rate:
300-700 ml/min in steps of 20 ml/min
Accuracy:
±10% or 50 ml/min whichever is largest
Degassing:
By use of negative pressure, -610 mmHg.
Adjustable degassing pressure between
-300 and -700 mmHg (-650 mmHg with
700 ml/min
fl
ow rate).
Accuracy:
±40 mmHg
Dialysis
fl
uid pressure:
-400 to +300 mmHg.
Accuracy:
±10 mmHg or ±5%
Proportioning of concen-
trates:
The proportioning of concentrate is
done through conductivity control. The
concentrates are pumped into the system
with one/two volume trically supervised
pumps. No minimum feeding pressure
is necessary, the
fl
uid is sucked in.
Maximum feeding pressure, 150 kPa.
Acetate:
Na
+
, 130 to 160 mmol/l, (±3 mmol/l)
Bicarbonate:
Na
+
, 130 to 160 mmol/l, (±6
mmol/l)HCO
3
- , 20 to 40 mmol/l,
(±6 mmol/l)
Measuring range:
13 to 16 mS/cm
Alarm limits:
±5% of the calculated conductivity set
value
HCEN9369 Revision.10.2005
Program version 12.xx
AK 95 S
Service
m anual - Technical data and spec i
fi
cations
3
:4
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...