6:3
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
Calibrations
CPU
Testpoint Ref
Tetspoint
Circuit board
Limits
(mV)
CPU-C
TP 6
TP 5
Control Driver board
45.00 - 55.00
CPU-P
TP 1
TP 2
Protective Driver board 45.00 - 55.00
CPU-W
TP 3
TP 4
Protective CPU board
4500 - 5500
MM 95
R 24
R 24
MM I/O board
45.00 - 55.00
AD Converters
1.
Start the AK 95 S, enter the Service menu
and select Calibrations.
2.
Choose
AD-converter.
When selecting “AD-Converter” from the display
with the SET knob and press
key following
display appears:
SELECT AD-CONVERTER
CPU-C
CPU-P
CPU-W
MM95
3.
Select an AD-converter with the SET knob
and press
key.
CPU X AD LOW
REF SET HIGH
xx.xx mV xx.xx xx.xx xx.xx
4.
Measure the voltage (using a 4.5 digit
voltmeter), between following testpoints or
across the resistor:
TP 6
TP 5
CPU C driver board
CPU C A/D ref
AD ref voltage range:
45.00 to 55.00 mV
Buzzer
TP2
Z0VAD
PVREF
TP1
AD ref voltage range:
45.00 to 55.00 mV
CPU W PROM
CPU W
SafeRAM
TP4
Z0VRW
TP3
WREF
AD ref voltage range:
4.500 to 5.500 V
R24
MM I/O AD ref voltage range:
45.00 to 55.00 mV
pH plug
(TP1)
(TP2)
5.
Set the “AD REF VOLTAGE” on the
display to the value shown on the voltme-
ter and press
.
6.
Press
START
UF
STOP
to load the value into the
EEPROM.
CPU X
AD-REF VOLTAGE STORED
7.
Press
to exit.
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...