4:24
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
Service Technicians Guide
UF diagnostics
DIAGNOSTIC FM MENU
Var flow Pumps UF
Valves
Select “UF” with the SET knob and press the
SELECT key and following display appears:
SELECT UF CELL TARATION
Complete
Select “Complete” with the SET knob and press
the SELECT key and following display appears:
COMPLETE UF CELL TARATION
Please wait
A complete taration as in treatment is now
performed. When ready following display occurs:
COMPLETE UF CELL TARATION READY
Valves diagnostics
DIAGNOSTIC FM MENU
Var flow Pumps UF
Valves
Select “Valves” with the SET knob and press the
SELECT key and following display appears:
INLET VALVE (INVA)
Press SELECT key to toggle valve
Turn the SET knob and all the different valves
in the machine that are possible to turn on and
off will “scroll” on the Information display.
When the SELECT key is pressed the selected
valve toggles on or off. When the valves not are
selected they are off.
INLET VALVE
INVA
AIR INLET VALVE
AIVA
RECIRCULATION VALVE
REVA
FLUSHING VALVE
FLVA
RINSING VALVE
RIVA
ZEROING VALVE
ZEVA
DIRECT VALVE CPU C
DIVA
DIRECT VALVE CPU P
DIVA
TARATION VALVE
TAVA
EVACUATION VALVE
EVVA
BYPASS VALVE CPU C
BYVA
BYPASS VALVE CPU P
BYVA
Note!
Be careful when open the AIVA to avoid
Fluid leakage into the monitor.
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...