Technical Data
3:8
AK 95 HCEN9135 Rev. 10.2005
Mains plug:
For 100/110V, Hospital grade, earth plug, type IEC 83 A5-
15
For 230V, Dual earth system plug, type IEC 83 C4
Maximum earth leakage
current:
100
µ
A (100V)
150
µ
A (110V)
205
µ
A (230V, 50 Hz )
400
µ
A (230V, 60 Hz )
Maximum patient leakage
current:
9
µ
A (100V)
11
µ
A (110V)
20
µ
A (230V)
Note:
All leakage currents specified are without external equipment
connected to the AK 95.
Connection of external equipment
External connector:
25 pin D-Sub with RS-232C or RS-422/Current loop.
Opto insulated fulfilling IEC 60601-1-1.
External equipment:
External equipment intended for connection to signal input,
signal output or other connectors, shall comply with relevant
IEC standard (e.g. IEC 60950 for IT equipment and the IEC
60601 series for medical electrical equipment). In addition, all
such combinations –
systems
– shall comply with the
standard IEC 60601-1-1,
Safety requirements for medical
electrical systems
. Equipment not complying to IEC 60601
shall be kept outside the patient environment, as defined in
the standard.
1)
Any person who connects external equipment to signal input,
signal output or other connectors has formed a system and is
therefore responsible for the system to comply with the
requirements of IEC 60601-1-1. If in doubt, contact qualified
technician or your local representative.
1)
The normal distance can be mentioned i.e. at least 1.5 m
from the patient or the patient support.
Extern alarm:
Max voltage:
24 V
AC or DC
Max current :
100 mA
AC or DC
RS-232C:
Max input voltage:
±15 V
DC
High level min output voltage:
5.0 V
DC
Low level max output voltage:
-5.0 V
DC
Max output current :
±5 mA
DC
RS-422/Current loop :
Typ current :
20 mA
DC
Max current :
50 mA
DC
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...