AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
6:6
Calibrations
Arterial Pressure Transducer Calibration
1.
Connect the Gambro reference instrument
(together with a calibration tube set) to
the arterial pressure transducer on the
front of the AK 95 S, see fig. below.
2.
Zeroset the reference instrument, wait for
stable reading.
3.
Start the AK 95 S, enter the
Service
menu and select
Calibrations
.
4.
Select
Transducers
,
External
and
ARTERIAL PRESSURE TRANSDUCER
from the selection diagram.
5.
Press
to confirm the selected trans-
ducer.
6.
Apply zero pressure by opening the three
clamps A, B and C on the calibration tube
set. Set the “REF SET” value on the
AK 95 to 0 mmHg.
7.
Press
to enter the setting.
8.
If the lit
START
UF
STOP
key is pressed at this time
the machine stores the
offset
for the
transducer in the EEPROM and updates
the checksum.
For the
sensitivity calibration
(the slope):
Close clamp A. Apply a pressure of approx.
-500 mmHg, shown on the reference in-
strument. Close clamp B.
ART PR
LOW
REF SET
HIGH
1.2
-550
-500
550
9.
Set the “REF SET” value on the AK 95 S
to the reading on the reference instrument
and press
twice.
10.
Open clamp A and B, then close clamp A
again. Apply a pressure of approx. +500
mmHg, shown on the reference instru-
ment. Close clamp B.
ART PR
LOW
REF SET
HIGH
1.2
-550
+500
550
11.
Set the “REF SET” value on the AK 95 S
to the reading on the reference instrument
and press
.
12.
Press
START
UF
STOP
to load the calibration values
into the EEPROM.
13.
When the EEPROM storage is ready,
press
to exit.
If a calibration error occurs:
•
Try to do the calibration again.
•
If the error occurs a second time:
Change the transducer and the transducer
board.
Report to your Gambro service representa-
tive the error message on the error dis-
play.
B
A
C
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...