AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
Technical Description
2:7
Treatment
Fluid part:
This technical description of the AK 95 S is
based on the flow diagram on page 2:5.
Water is taken in through a pressure regulator,
PR 1
,
and a
heat exchanger
, before entering
the AK 95 S fluid-unit. The pressure regulator is
set to approximately 600 mmHg, measured after
the heat exchanger.
The inlet pressure switch,
INPS
, detects too
low pressure. If the inlet pressure is too low (less
than +99 mmHg), the inlet valve,
INVA
, is
closed, the heater is switched off and the
AK 95 S enters the fluid bypass mode. An atten-
tion will be given when the pressure is too low.
The inlet pressure should be constant, but
normally there are small fluctuations. The
second pressure regulator,
PR 2
, is used to
ensure a constant pressure, approximately
+130 mmHg. Normaly during treatment INVA
is open.
The main function of the temperature transduc-
ers in the AK 95 S is to measure the dialysis
fluid temperature and control the heat regula-
tion. The regulating temperature transducer,
T
Reg
,
is located after the heater.
The T
Reg
transducer is used in the fast tempera-
ture control loop. The set value for the loop is the
set value entered by the user adjusted for the
measured temperature drop to the conductivity
cell B.
The temperature drop is calculated when the
temperature measured at cond.cell B is stable.
The temperature considers to be stable when it
has been in a window of ± 0.25 °C for one minute
or more.
The function of the
flow switch
is to sense if
there is fluid flow or not through the heater.
If the flow switch does not detect flow, the
heater will be switched off in order to protect the
heater from overheating.
There is a temperature transducer in each
conductivity cell, used for compensating the
conductivity calculation.
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...