Technical Data
3:6
AK 95 HCEN9135 Rev. 10.2005
Hypochlorite Program (presetable)
Concentration of disinfectant
10 % available chlorine
Concentration in machine
0.5 %; i.e. diluted 1 + 19 (1:20)
Volume
Approx. 145 ml (with UFD-kit approx. 153 ml)
Contact time between treatments
10 minutes
Contact time overnight or when not in
use
Maximum 20 min, not intended for
overnight disinfection!
Total time
30 minutes (with UFD-kit 48 minutes)
including 10 min contact time
Formaldehyde Program (presetable)
Concentration of disinfectant
37 % formaldehyde
Concentration in machine
3.7 %; i.e. diluted 1 + 9 (1:10)
Volume
Approx. 275 ml (with UFD-kit approx. 290 ml)
Contact time between treatments
20 min
Contact time overnight or when not in
use (recommended with UFD-kit)
Minimum 6h dwell time
Total time
119 minutes (with UFD-kit 150 minutes)
including 20 min contact time
Heat disinfection:
One of three alternatives for heat disinfection can be selected.
The second alternative is presetable for a combined heating
program and the third is a CleanCart / heating alternative. The
default settings are as follows.
Total Time [min]
AK 95
Disinfection programs
230 V
110 V
100 V
Heat
35
39
49
With
UFD
installed
Heat Citric acid 20%
52
57
67
Heat CleanCart
49
54
64
Heat
33
37
47
Without UFD
Heat Citric acid 20%
50
55
65
Heat CleanCart
46
51
61
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...