5:8
Maintenance Manual
AK 95 / AK 95 S
HCEN12189 Rev.10
. 2006
Actions to carry out after the parts
have been exchanged
Base-kit
1.
Check that the screws for the blood pump
are tightened. (Both inside the BM and
behind the blood pump rotor)
2.
Check that the screws for the hose clamps
on the water inlet and drain tube are
tightened.
3.
Check that the wheels are tightened.
4.
Perform the Protective Earth Test (PET)
according to instructions available in
HCEN12190, Electrical safety inspections
for AK-machines.
5.
Check the protective earth resistance:
Measure between TP 5 (Z0VL) on the
control CPU board and the metal on the
heater rod.
Measure in
M
Ω
,
should be > 1 M
Ω
. The
machine must be empty and switched off
during measurement.
Note!
Mount the blood monitor on the
fluid monitor.
6.
Erase the error code buffers. Enter service
menu > Diagnose > System > Errors >
Erase.
7.
Calibrate the A/D converters for AK 95 S,
see chapter "Calibrations" in the Service
manual.
8. BPM-test
Note!
Following text is valid for machines equipped
with the SunTech blood pressure module
(K24274001). For machines equipped with the
Colin blood pressure module (K23564001), refer
to HCEN9095, revision 10.2005 for technical and
maintenance information about BPM.
8.1 Test method:
-
Attach BPM cuff, 100900141, tight to the
BPM
test equipment, K22151001.
-
Start the AK 95 S with the main switch
and the ON/OFF button.
-
Enter Service mode.
-
Enter Diagnostics mode.
-
Select BM, Others, BPM
8.2 BPM sub-tests
In the BPM service menu the following sub-tests
are included:
- Air leakage test: measures that the equipment
is airtight.
- Inflation speed test: measures the time to fill
the cuff with air.
- Static pressure: Check of pressure transducer
calibration
Each of the tests can be carried out separately
but the above order for carrying out the sub-tests
is recommended.
Check that the results from the sub-tests not
exceed following limits:
- Air leakage: maximum 18 mmHg/ 3 min
- Inflation speed: 2 -11 s
- Static pressure: Maximum deviation from
reference to be within ± 3 mm Hg.
50 ± 3 mmHg
150 ± 3 mmHg
250 ± 3 mmHg
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...