5:10
Maintenance Manual
AK 95 / AK 95 S
HCEN12189 Rev.10
. 2006
Base-kit
9.
Priming detector calibration
9.1
Start the AK 95 S, enter the Service menu
and select Calibrations.
9.2
Select Transducer, External and PRIMING
DETECTOR from the selection diagram.
9.3
Press "Select" to start the calibration
routine.
9.4
Insert an empty tube in the priming
detector and press "Select".
Use the normal tube from the bloodline.
9.5
The calibration is now done automatically
and the calibration factor is written into
the EEPROM.
9.6
When the info. display shows "EEPROM
STORAGE COMPLETED", the calibration
is completed.
9.7
Press "Hold" to exit.
Note!
If a technical error is detected, the
calibration must be restarted, or the
priming detector replaced.
10.
Measure the power on battery B1 on the
Protective driver board, exchange if the
value is below 2,6 V (
Note!
The machine
must be switched on when measuring).
11.
Control of the blood pump occlusion
Check surface on the roller at the pump
rotor, if it is damaged exchange the unit.
The roller unit should easily go back to
upper end position when it has been
pressed down to end position, exchange
unit if it gets stuck in any position.
Note!
If the blood lines have a different
wall thickness, the occlusion of the blood
pump must be adjusted to the right size.
Use the gauge pins (stop/go) available in
the kit K40158001. The kit includes
following three different gauge pins:
T = 1.1 (Stop pin = 1.1*2*0.7 = 1.5mm)
T = 1.6 (Stop pin = 1.6*2*0.7 = 2.2mm)
T = 2.0 (Stop pin = 2.0*2*0.7 = 2.8mm)
T = wall thickness of the blood tubing
Use following formula to calculate which
gage pin to use:
2 * wall thickness * 0,7 (mm)
Check the range between the blood pump
rollers and the path according to the
picture below.
Adjust range between the blood pump
rollers and the path if it is necessary.
The GO pin should barely pass within the
calibrating area.
The STOP pin is NOT allowed to pass in
any point within the calibrating area.
Note!
Put grease on the GO and STOP
pins when not used and store gage in
plastic bag to prevent it from corroding.
Note!
The gauge pins can also be ordered
separately:
T=1.1 (white) K40158002
T=1.6 mm (red) K40158003
T=2.0 mm (blue) K40158004
= Measuring
Summary of Contents for AK 95
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Page 5: ...This page is intentionally left blank ...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Page 247: ...AK 95 AK 95 S Service Manual ...